JVDI Advertisement
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


Journal of Veterinary Diagnostic Investigation Vol. 20 Issue 1, 83-85
Copyright © 2008 by the American Association of Veterinary Laboratory Diagnosticians
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by van Oldruitenborgh-Oosterbaan, M. M. S.
Right arrow Articles by Spierenburg, A. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by van Oldruitenborgh-Oosterbaan, M. M. S.
Right arrow Articles by Spierenburg, A. J.

Brief Communications

Evaluation of the usefulness of the portable device Lactate Pro for measurement of lactate concentrations in equine whole blood

Marianne M. Sloet van Oldruitenborgh-Oosterbaan1, Erica T. W. van den Broek and Astrid J. Spierenburg

Correspondence: 1Corresponding Author: Marianne M.S. van Oldruitenborgh-Oosterbaan, Department of Equine Sciences, Faculty of Veterinary Medicine, University of Utrecht, the Netherlands, Yalelaan 114, 3584 CM Utrecht, e-mail: m.sloet{at}vet.uu.nl

Blood lactate measurements are commonly used in exercising horses to determine the onset of lactate accumulation and in colic patients to assess clinical status and to indicate prognosis. To study the usability of a portable blood lactate meter based on dry chemistry (Lactate Pro), the data from this instrument were compared to data from a laboratory-used lactate meter based on wet chemistry (ABL 605 blood gas analyzer [ABL]). Heparinized blood samples were obtained from horses participating in a jumping experiment (n = 9), from horses cantering at maximal speed on a racetrack (n = 7), and from patients admitted to the Department of Equine Sciences at Utrecht University for severe colic (n = 13). Seventeen of these samples were tested in duplicate on both instruments to determine the repeatability of the measurements. Blood lactate concentrations measured with the Lactate Pro ranged from 0.8–17.6 mmol/liter and with the ABL from 1.0–18.6 mmol/liter. The correlation between lactate concentrations obtained using the Lactate Pro and values from the ABL was 0.90, and the relationship was represented by the following formula: y = 0.90 · x + 0.36, indicating a linear relationship between values produced by the ABL and Lactate Pro. The repeatability for the Lactate Pro was high (0.997), which is comparable to the ABL (0.999).

Key Words: Horses • lactate measurements • Lactate Pro